Subwavelength registration of skull-compensating acoustic holograms via nonlinear parametric array feedback
Source: arxiv.org/abs/2611.15730
Realizing the potential of optimized acoustic holograms for applications like transcranial ultrasound requires sub-wavelength alignment accuracy (less than 1.5 mm at 1 MHz), which is difficult to achieve with current non-MRI optical neuro-navigation methods. To address the critical challenge of alignment, we leverage the parametric array (PA) signal as an alignment surrogate, a highly penetrative, low-frequency difference wave generated by the nonlinear mixing of two primary high-frequency beams. Simulations show that a misaligned skull-compensating lens increases the primary acoustic energy…